Manufacturer: TestEquity
Model: 1027S
Reference No.: 280LC
Description: TestEquity Model 1027S Temperature Chamber
Model 1027S Temperature Chamber Specifications:
Temperature Range
-35°C to +175°C
Control Tolerance
±0.5°C (Short-term variations measured at the control sensor after stabilization)
Uniformity
±1.0°C (Variations throughout the chamber after stabilization)
Cool Down Transition Time (empty)* |
Start Temp |
End Temp |
+23°C |
0°C |
-10°C |
-20°C |
-30°C |
-35°C |
+150°C |
9.0 min |
14 min |
17 min |
20 min |
26 min |
Ultimate |
+85°C |
4.0 min |
7.5 min |
10 min |
13 min |
19 min |
+23°C |
—– |
1 min |
2 min |
4 min |
9 min |
Cool Down Transition Time (with 80 lb. aluminum load)* |
Start Temp |
End Temp |
+23°C |
0°C |
-10°C |
-20°C |
-30°C |
-35°C |
+150°C |
20.0 min |
28 min |
33 min |
38 min |
46 min |
Ultimate |
+85°C |
8.0 min |
16 min |
20 min |
25 min |
33 min |
Heat Up Transition Time (empty)* |
Start Temp |
End Temp |
+23°C |
+50°C |
+85°C |
+125°C |
+150°C |
+175°C |
+23°C |
—– |
1.5 min |
7 min |
14 min |
20 min |
25 min |
0°C |
1.5 min |
3.5 min |
13 min |
20 min |
23 min |
31 min |
-10°C |
2.2 min |
4.2 min |
14 min |
22 min |
25 min |
33 min |
-35°C |
5.5 min |
10 min |
16 min |
23 min |
29 min |
34 min |
Heat Up Transition Time (with 80 lb. aluminum load)* |
Start Temp |
End Temp |
+23°C |
+50°C |
+85°C |
+125°C |
+150°C |
+175°C |
0°C |
5 min |
13 min |
23 min |
36 min |
45 min |
55 min |
-35°C |
11 min |
19 min |
29 min |
42 min |
51 min |
61 min |
Rate Of Change
To calculate rate of change for a particular condition, take the difference between the Start Temp and End Temp and divide by the Transition Time.
Cool Down Example (empty): From +85°C to 0°C = 85°C / 7.5 min = 11.3°C/min.
Cool Down Example (with 80 lb. load): From +85°C to 0°C = 85°C / 16 min = 5.3°C/min.
Heat Up Example (empty): From 0°C to +85°C = 85°C / 13 min = 6.5°C/min.
Heat Up Example (with 80 lb. load): From 0°C to +85°C = 85°C / 23 min = 3.7°C/min.
*Note: Transition times are measured after a 2 hour soak at the respective start temperature with an empty chamber, as indicated on the temperature controller, 23°C ambient. Measured with setpoint beyond the start and end temperatures. Does not include the effect of proportional band when approaching setpoint. Performance is reduced by 17% with 50 Hz input power.
Live Load Capacity
+23°C |
0°C |
-10°C |
-20°C |
-30°C |
4,000 Watts |
2,800 Watts |
2,300 Watts |
1,800 Watts |
1,250 Watts |
Refrigeration and Heating System
Refrigerant |
R-404A (Dupont HP-62) |
Compressors |
3.5 HP Copeland scroll compressor. More about Scroll Compressors >> |
Condenser |
Air Cooled |
Heat of Rejection |
30,200 BTUH (maximum rated chamber load at maximum cooling rate from high temperature soak) |
Heater Power |
4,000 Watts @ 208 V input |
Air Flow |
830 cfm |
Instrumentation
Temperature Controller |
Watlow F4T Touch Screen Controller with RS-232, Ethernet interface, 4.3″ color graphic touch screen. OR…
Watlow F4 Controller with RS-232 interface, LED readout of temperature, LCD display of other parameters. |
Limit Controller |
Independent high and low temperature limits. Triggers an audible alarm and shuts down the chamber. Relay contacts provide a safety power interlock for test sample. |
Chart Recorder |
(Optional) Honeywell DR4300 Series. One pen, 10″ circular chart. Mounts in lower front door. |
 |
Input Power Requirements
230 V ±10%, 60 Hz, 3 Phase |
Max Current Draw 31 A; Recommended Service 40 A |
208 V -5/+10%, 60 Hz, 3 Phase |
Max Current Draw 26 A; Recommended Service 35 A |
|
Input may be configured for 230 V or 208 V in the field by changing jumpers. Three phase load is balanced. Call for other voltages or 50 Hz operation.
Customer power source must be hard-wired to the chamber by a qualified electrician. Power cord and plug is not included. |
Physical Characteristics and Safety
Inside Dimensions |
40″ W x 32″ H x 36.5″ D (27 cubic feet)
1016 mm W x 812.8 mm H x 927 mm D (764 liters) |
Outside Dimensions |
49″ W x 73.25″ H x 63″ D (nominal)
1244 mm W x 1860 mm H x 1600 mm D
Door latch adds 3″ to width on right side. Circulator motor and housing adds 6″ to height – may be removed for move-in. |
 |
Minimum Installed Clearance |
18″ from the left and right side
24″ from the rear |
Window Viewing Area |
16″ W x 13″ H |
Access Ports |
4″ Port on left and right side (two total)
Supplied with foam plugs |
Weight |
Chamber Weight: 1,410 pounds
Shipping Weight: 1,595 pounds |
NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. Designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result in decreased cooling performance. Additional ports and shelves will also affect performance. Operation above 30°C (85°F) or below 16°C (60°F) ambient is not recommended. Performance is reduced by 17% with 50 Hz input power.
Due to continuous product improvements, specifications subject to change without notice.
Contact SMT Sales Group for additional details.
Condition: Complete & Operational
Availability: Immediate for purchase / 2-4 weeks for shipping
Location & Shipping: USA / FOB Origin
Actual Machine Photos:
Contact SMT Sales Group using the form below or email
sales@smtsalesgroup.com for additional information or inquiries.
FAST response, within an hour or less.